{"data":{"id":"us-va/12vac5-481-2048","jurisdiction":"us-va","citation":"12VAC5-481-2048","heading":"Full-inspection servicing for teletherapy and gamma stereotactic radiosurgery units","body":"A. Licensees shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully inspected and serviced during each source replacement to assure proper functioning of the source exposure mechanism and other safety components. The interval between each full-inspection servicing shall not exceed five years for each teletherapy unit and shall not exceed seven years for each gamma stereotactic radiosurgery unit.\n\nB. This inspection and servicing may only be performed by person specifically licensed to do so by the agency, the NRC, or another Agreement state.\n\nC. Licensees shall keep a record of the inspection and servicing in accordance with 12VAC5-481-2070 X.","path":["Title 12. Health","Agency 5. Department of Health","Chapter 481. Virginia Radiation Protection Regulations","Part VII. Use of Radionuclides in the Healing Arts","Article 9. Photon Emitting Remote Afterloader Units, Teletherapy Units, and Stereotactic Radiosurgery Units"],"source_url":"https://law.lis.virginia.gov/admincode/title12/agency5/chapter481/section2048/","current_through":"2026 Regular Session (effective July 1, 2026)","vintage":"","retrieved_at":"2026-09-14T04:50:29Z","sha256":"b8bd804d37a2e45337e55acc960e51dd50f7926f6127140765d375d14dd00a19","source_id":"us-va-vac","stale":false,"prev":"us-va/12vac5-481-2047","next":"us-va/12vac5-481-2049"},"notice":"GroundRules: Original legal text. Not legal advice."}