{"data":{"id":"us/10-cfr-35.655","jurisdiction":"us","citation":"10 CFR 35.655","heading":"Full-inspection servicing for teletherapy and gamma stereotactic radiosurgery units.","body":"(a) A licensee shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully inspected and serviced during each source replacement to assure proper functioning of the source exposure mechanism and other safety components. The interval between each full-inspection servicing shall not exceed 5 years for each teletherapy unit and shall not exceed 7 years for each gamma stereotactic radiosurgery unit.\n(b) This inspection and servicing may only be performed by persons specifically licensed to do so by the Commission or an Agreement State.\n(c) A licensee shall keep a record of the inspection and servicing in accordance with § 35.2655.","path":["Title 10—Energy","CHAPTER I—NUCLEAR REGULATORY COMMISSION","PART 35—MEDICAL USE OF BYPRODUCT MATERIAL","Subpart H—Photon Emitting Remote Afterloader Units, Teletherapy Units, and Gamma Stereotactic Radiosurgery Units"],"source_url":"https://www.ecfr.gov/api/versioner/v1/full/2026-08-25/title-10.xml","current_through":"2026-08-25","vintage":"","retrieved_at":"2026-08-27T02:24:09Z","sha256":"da29f68fdcbe2e9cfada90c73cc5bf8adba9d02b1feeb303a0ed0edb95045663","source_id":"us-cfr","stale":true,"prev":"us/10-cfr-35.652","next":"us/10-cfr-35.657"},"notice":"GroundRules: Original legal text. Not legal advice."}