{"data":{"id":"us/47-cfr-95.2369","jurisdiction":"us","citation":"47 CFR 95.2369","heading":"WMTS field strength limits.","body":"Each WMTS transmitter type must satisfy the field strength limits in this section.\n(a) For WMTS transmitter types operating in the 608-614 MHz band, the field strength of the transmitted signal must not exceed 200 mV/m, measured at a distance of 3 meters, using instrumentation with a CISPR quasi-peak detector.\n(b) For WMTS transmitter types operating in the 1395-1400 MHz and 1427-1432 MHz bands, the field strength of the transmitted signal must not exceed 740 mV/m, measured at 3 meters, using instrumentation with an averaging detector and a 1 MHz reference bandwidth.","path":["Title 47—Telecommunication","CHAPTER I—FEDERAL COMMUNICATIONS COMMISSION","SUBCHAPTER D—SAFETY AND SPECIAL RADIO SERVICES","PART 95—PERSONAL RADIO SERVICES","Subpart H—Wireless Medical Telemetry Service"],"source_url":"https://www.ecfr.gov/api/versioner/v1/full/2026-08-25/title-47.xml","current_through":"2026-08-25","vintage":"","retrieved_at":"2026-08-27T02:26:28Z","sha256":"003c00c90fe15f037205ae99b888529ddf8726d94a5fd4376e8c885b3c2f2965","source_id":"us-cfr","stale":true,"prev":"us/47-cfr-95.2367","next":"us/47-cfr-95.2371-95.2377"},"notice":"GroundRules: Original legal text. Not legal advice."}