{"data":{"id":"us/47-cfr-95.2379","jurisdiction":"us","citation":"47 CFR 95.2379","heading":"WMTS unwanted emissions limits.","body":"Each WMTS transmitter type must be designed to comply with the requirements in this paragraph.\n(a) Unwanted emissions on frequencies below 960 MHz must not exceed 200 μ V/m, measured at a distance of 3 meters using measuring instrumentation with a CISPR quasi-peak detector.\n(b) Unwanted emissions on frequencies above 960 MHz must not exceed 500 μ V/m, measured at a distance of 3 meters using measuring equipment with an averaging detector and a 1 MHz measurement bandwidth.","path":["Title 47—Telecommunication","CHAPTER I—FEDERAL COMMUNICATIONS COMMISSION","SUBCHAPTER D—SAFETY AND SPECIAL RADIO SERVICES","PART 95—PERSONAL RADIO SERVICES","Subpart H—Wireless Medical Telemetry Service"],"source_url":"https://www.ecfr.gov/api/versioner/v1/full/2026-08-25/title-47.xml","current_through":"2026-08-25","vintage":"","retrieved_at":"2026-08-27T02:26:28Z","sha256":"c518335c93a3a6c801cd5be76b6b788781dc9f6a771b82abea44c935d8db32af","source_id":"us-cfr","stale":true,"prev":"us/47-cfr-95.2371-95.2377","next":"us/47-cfr-95.2381-95.2383"},"notice":"GroundRules: Original legal text. Not legal advice."}