{"data":{"id":"us/47-cfr-95.2565","jurisdiction":"us","citation":"47 CFR 95.2565","heading":"MedRadio frequency accuracy.","body":"Each MedRadio transmitter type must be designed to maintain a frequency stability of ±100 ppm of the operating frequency over the applicable temperature range set forth in this section. Frequency stability testing shall be performed over the appropriate temperature range.\n(a) 25 °C to 45 °C in the case of medical implant transmitters; and\n(b) 0 °C to 55 °C in the case of MedRadio programmer/control transmitters and medical body-worn transmitters.","path":["Title 47—Telecommunication","CHAPTER I—FEDERAL COMMUNICATIONS COMMISSION","SUBCHAPTER D—SAFETY AND SPECIAL RADIO SERVICES","PART 95—PERSONAL RADIO SERVICES","Subpart I—Medical Device Radio Communications Service"],"source_url":"https://www.ecfr.gov/api/versioner/v1/full/2026-08-25/title-47.xml","current_through":"2026-08-25","vintage":"","retrieved_at":"2026-08-27T02:26:28Z","sha256":"f37d1f6abb7b6d433d30527254c2da16b00c8d631bb843ff20c215a6305b75ed","source_id":"us-cfr","stale":true,"prev":"us/47-cfr-95.2563","next":"us/47-cfr-95.2567"},"notice":"GroundRules: Original legal text. Not legal advice."}