{"data":{"id":"us/47-cfr-95.2567","jurisdiction":"us","citation":"47 CFR 95.2567","heading":"MedRadio radiated power limits.","body":"Each MedRadio transmitter type must be designed such that the MedRadio equivalent isotropically radiated power (M-EIRP) does not exceed the limits in this section. Compliance with these limits must be determined as set forth in § 95.2569.\n(a) Transmitters subject to frequency monitoring—401-406 MHz. For MedRadio transmitters that are not excepted under § 95.2559(b) from the frequency monitoring requirements of § 95.2559(a):\n(1) The M-EIRP within any 300 kHz bandwidth within the 402-405 MHz band must not exceed 25 microwatts.\n(2) The M-EIRP within any 100 kHz bandwidth within the 401-402 MHz or 405-406 MHz bands must not exceed 25 microwatts.\n(b) Transmitters excepted from frequency monitoring—401-402 MHz and 405-406 MHz. For MedRadio transmitters that are excepted under § 95.2559(b)(2) or (3) from the frequency monitoring requirements of § 95.2559(a):\n(1) The M-EIRP of any transmitter operating in the 401-401.85 MHz or 405-406 MHz bands must not exceed 250 nanowatts in any 100 kHz bandwidth.\n(2) The M-EIRP of any transmitter operating in the 401.85-402 MHz band must not exceed 25 microwatts in any 150 kHz bandwidth.\n(c) Transmitters excepted from frequency monitoring—403.65 MHz. For MedRadio transmitters that are excepted under § 95.2559(b)(4) from the frequency monitoring requirements of § 95.2559(a), the M-EIRP must not exceed 100 nanowatts in the 300 kHz bandwidth centered at 403.65 MHz.\n(d) Transmitters—other frequency bands. For MedRadio transmitters operating in the 413-419 MHz, 426-432 MHz, 438-444 MHz, or 451-457 MHz bands:\n(1) The peak M-EIRP over the frequency bands of operation must not exceed the lesser of zero dBm (1 mW) or 10 log (B)−7.782 dBm, where B is the MedRadio 20 dB emission bandwidth in megahertz.\n(2) The peak power spectral density must not exceed 800 microwatts per megahertz in any one megahertz band.\n(e) Transmitters—2360-2390 MHz band. For MedRadio transmitters operating in the 2360-2390 MHz band, the M-EIRP over the bands of operation must not exceed the lesser of zero dBm (1 mW) or 10 log (B) dBm, where B is the MedRadio 20 dB emission bandwidth in megahertz.\n(f) Transmitters—2390-2400 MHz band. For MedRadio transmitters operating in the 2390-2400 MHz band, the M-EIRP over the bands of operation must not exceed the lesser of 13 dBm (20 mW) or 16 + 10 log (B) dBm, where B is the MedRadio 20 dB emission bandwidth in megahertz.","path":["Title 47—Telecommunication","CHAPTER I—FEDERAL COMMUNICATIONS COMMISSION","SUBCHAPTER D—SAFETY AND SPECIAL RADIO SERVICES","PART 95—PERSONAL RADIO SERVICES","Subpart I—Medical Device Radio Communications Service"],"source_url":"https://www.ecfr.gov/api/versioner/v1/full/2026-08-25/title-47.xml","current_through":"2026-08-25","vintage":"","retrieved_at":"2026-08-27T02:26:28Z","sha256":"f1f08d0b5cfb0fddbf560be84c190aefeb56099fc20c2c1ea8660fcc850f2076","source_id":"us-cfr","stale":true,"prev":"us/47-cfr-95.2565","next":"us/47-cfr-95.2569"},"notice":"GroundRules: Original legal text. Not legal advice."}