12VAC5-481-2017: Therapy-related computer systems
Where this section sits in the code
- Title 12. Health
- Agency 5. Department of Health
- Chapter 481. Virginia Radiation Protection Regulations
- Part VII. Use of Radionuclides in the Healing Arts
- Article 7. Manual Brachytherapy
Licensees shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing shall include, as applicable, verification of:
1. The source-specific input parameters required by the dose calculation algorithm;
2. The accuracy of dose, dwell time, and treatment time calculations at representative points;
3. The accuracy of isodose plots and graphic displays; and
4. The accuracy of the software used to determine sealed source positions from radiographic images.
Collected 2026-09-14T04:50:29Z. Source file · JSON