21 CFR 803.56: If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?

Read at publisher ↗
Where this section sits in the code
  1. Title 21—Food and Drugs
  2. CHAPTER I—FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES
  3. SUBCHAPTER H—MEDICAL DEVICES
  4. PART 803—MEDICAL DEVICE REPORTING
  5. Subpart E—Manufacturer Reporting Requirements

If you are a manufacturer, when you obtain information required under this part that you did not provide because it was not known or was not available when you submitted the initial report, you must submit the supplemental information to us within 30 calendar days of the day that you receive this information. You must submit the supplemental or followup report in accordance with the requirements of § 803.12(a). On a supplemental or followup report, you must:

(a) Indicate that the report being submitted is a supplemental or followup report;

(b) Submit the appropriate identification numbers of the report that you are updating with the supplemental information (e.g., your original manufacturer report number and the user facility or importer report number of any report on which your report was based), if applicable; and

(c) Include only the new, changed, or corrected information.

Collected 2026-08-27T02:24:47Z. Source file · JSON