21 CFR 882.1935: Near Infrared (NIR) Brain Hematoma Detector.
Where this section sits in the code
- Title 21—Food and Drugs
- CHAPTER I—FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES
- SUBCHAPTER H—MEDICAL DEVICES
- PART 882—NEUROLOGICAL DEVICES
- Subpart B—Neurological Diagnostic Devices
(a) Identification. A Near Infrared (NIR) Brain Hematoma Detector is a noninvasive device that employs near-infrared spectroscopy that is intended to be used to evaluate suspected brain hematomas.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) The sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) The labeling must include specific instructions and the clinical training needed for the safe use of this device;
(3) Appropriate analysis/testing should validate electromagnetic compatibility (EMC), electrical safety, and battery characteristics;
(4) Performance data should validate accuracy and precision and safety features;
(5) Any elements of the device that may contact the patient should be demonstrated to be biocompatible; and,
(6) Appropriate software verification, validation, and hazard analysis should be performed.
Collected 2026-08-27T02:24:47Z. Source file · JSON