GroundRules
← Search the law
Federal regulations · Through 2026-08-25 · Newer source version available

42 CFR 84.153: Airflow resistance test, Type A and Type AE supplied-air respirators; minimum requirements.

Read at publisher ↗
Where this section sits in the code
  1. Title 42—Public Health
  2. CHAPTER I—PUBLIC HEALTH SERVICE, DEPARTMENT OF HEALTH AND HUMAN SERVICES
  3. SUBCHAPTER G—OCCUPATIONAL SAFETY AND HEALTH RESEARCH AND RELATED ACTIVITIES
  4. PART 84—APPROVAL OF RESPIRATORY PROTECTIVE DEVICES
  5. Subpart J—Supplied-Air Respirators

(a) Airflow resistance will be determined when the respirator is completely assembled with the respiratory-inlet covering, the air-supply device, and the maximum length of air-supply hose coiled for one-half its length in loops 1.5 to 2.1 m. (5 to 7 feet) in diameter.

(b) The inhalation resistance, drawn at the rate of 85 liters (3 cubic feet) per minute when the blower is not operating or under any practical condition of blower operation shall not exceed the following amounts:

Maximum length of hose for which respirator is approved Maximum resistance, water column height

Feet Meters Inches Millimeters

75 23 1.5 38

150 46 2.5 64

250 76 3.5 89

300 91 4.0 102

(c) The exhalation resistance shall not exceed 25 mm. (1 inch) of water-column height at a flow rate of 85 liters (3 cubic feet) per minute when the blower is not operating or under any practical condition of blower operation.

Collected 2026-08-27T02:26:11Z. Source file · JSON