42 CFR 84.204: Exhalation valve leakage test; minimum requirements.

Read at publisher ↗
Where this section sits in the code
  1. Title 42—Public Health
  2. CHAPTER I—PUBLIC HEALTH SERVICE, DEPARTMENT OF HEALTH AND HUMAN SERVICES
  3. SUBCHAPTER G—OCCUPATIONAL SAFETY AND HEALTH RESEARCH AND RELATED ACTIVITIES
  4. PART 84—APPROVAL OF RESPIRATORY PROTECTIVE DEVICES
  5. Subpart L—Chemical Cartridge Respirators

(a) Dry exhalation valves and valve seats will be subjected to a suction of 25 mm. water-column height while in a normal operating position.

(b) Leakage between the valve and valve seat shall not exceed 30 milliliters per minute.