47 CFR 95.2379: WMTS unwanted emissions limits.
Where this section sits in the code
- Title 47—Telecommunication
- CHAPTER I—FEDERAL COMMUNICATIONS COMMISSION
- SUBCHAPTER D—SAFETY AND SPECIAL RADIO SERVICES
- PART 95—PERSONAL RADIO SERVICES
- Subpart H—Wireless Medical Telemetry Service
Each WMTS transmitter type must be designed to comply with the requirements in this paragraph.
(a) Unwanted emissions on frequencies below 960 MHz must not exceed 200 μ V/m, measured at a distance of 3 meters using measuring instrumentation with a CISPR quasi-peak detector.
(b) Unwanted emissions on frequencies above 960 MHz must not exceed 500 μ V/m, measured at a distance of 3 meters using measuring equipment with an averaging detector and a 1 MHz measurement bandwidth.
Collected 2026-08-27T02:26:28Z. Source file · JSON