47 CFR 95.2565: MedRadio frequency accuracy.
Where this section sits in the code
- Title 47—Telecommunication
- CHAPTER I—FEDERAL COMMUNICATIONS COMMISSION
- SUBCHAPTER D—SAFETY AND SPECIAL RADIO SERVICES
- PART 95—PERSONAL RADIO SERVICES
- Subpart I—Medical Device Radio Communications Service
Each MedRadio transmitter type must be designed to maintain a frequency stability of ±100 ppm of the operating frequency over the applicable temperature range set forth in this section. Frequency stability testing shall be performed over the appropriate temperature range.
(a) 25 °C to 45 °C in the case of medical implant transmitters; and
(b) 0 °C to 55 °C in the case of MedRadio programmer/control transmitters and medical body-worn transmitters.